Reports until 11:29, Tuesday 22 April 2025
H1 AOS (SEI)
jason.oberling@LIGO.ORG - posted 11:29, Tuesday 22 April 2025 (84049)
Another Look at WHAM1 ISI Alignment

J. Oberling, J. Warner

After letting things settle overnight we took another look at the WHAM1 ISI alignment this morning.  Jim found that a few of the HEPI stops had become engaged (likely during the final moves yesterday afternoon), so he backed those off and I took a look at ISI alignment.  Again, the ISI position is reported as deviation from its nominal LHO global coordinates of [-22726.7, 0.0, -201.9] mm.

Jim said this was good enough to begin attaching HEPI actuators, so that work has begun.  We'll take another look at ISI alignment once the actuators are attached.

Data

Data for the above alignment deviations.  Method described in alog 84057.

X axis position and yaw:

Y axis position:

Z axis position and level: