Displaying report 1-1 of 1.
Reports until 18:43, Tuesday 22 April 2025
H1 CAL
anthony.sanchez@LIGO.ORG - posted 18:43, Tuesday 22 April 2025 (84075)
PCAL TX Module Maintence @ EX

Rick and I went down to End X to try and tune up the X End TX module. DCC T1600436-V12
We found the OFS offset to be set at 3.85V before we arrived.
A number of our measurements were made with the OFS Offset set to 7.7 V which is 100% of our useable OFS functionality. All of our low power measurments were done this way,  and perhaps we should have dialed that back to 95%  which may explain why some of our number look a little higher than last year's.
The Integrating sphere Apature target was put on the RX sensor.
The AOM position was adjusted, but resulted in no noticable difference in power.
The Beam spot on the TX PD was touched up, as it was not centered.
The Temperature and current for the Pumping Diode on the CrystaLaser Power supply was touched up to try and get some more performance out of it. 
And the spot on the internal spectralon shell of the RX sensor was blown out with some compressed air. Spot is now gone.
 

Date April-22-2025
Laser Shutter Check Pass
Max OFS Offset 7.7V
95% OFS Offset 7.315
Operating OFS Offset 3.6575
Laser Output Power 1.7
After-Laser Rejected Power                      34.6mW
AOM Input Power 1.65W
Max Diffracted Power 1.14W
Un-Diffracted Power                        0.307W
AOM Diffraction Efficiency 0.69
After-AOM Rejected Power                      15.2mW
TxPD Power                      11.9mW
OFSPD Power                        5.7mW
Outer Beam Power 0.536
Inner Beam Power 0.526
Output Beam Power Ratio 0.9813
OFS Gain 39.6
OFS Phase Margin 42.1
   
Images attached to this report
Non-image files attached to this report
Displaying report 1-1 of 1.