david.barker@LIGO.ORG - posted 07:24, Friday 02 May 2025 (84232)
Optics Lab dust monitor EPICS IOC restarted following crash

Last night at 23:35 Thu 01may2025 PDT the lab dust monitor EPICS IOC stopped running, causing EDC disconnected channels.

I logged into the IOC this morning, it was reporting serial comms issues with the LAB2 and LAB3 dust monitors, but I think that is expected (logs shown below). No logs seen about LAB1 or the IOC itself.

I restarted the IOC which cleared the EDC disconnections. CS_DUST_LAB1 is sampling again, albeit currently with zero particle counts. LAB2 and LAB3 are shown as "NOT IN USE".

 

2025/05/02 07:09:15.676960 gt521s1 H1:PEM-CS_DUST_LAB2_HOLDTIMEMON: No reply from device within 1000 ms
2025/05/02 07:09:16.830460 gt521s1 H1:PEM-CS_DUST_LAB2_300NM_RAW: No reply from device within 1000 ms
2025/05/02 07:09:17.933145 gt521s1 H1:PEM-CS_DUST_LAB2_STATE: No reply from device within 1000 ms
2025/05/02 07:09:19.034865 gt521s1 H1:PEM-CS_DUST_LAB3_STATE: No reply from device within 1000 ms
2025/05/02 07:09:20.138060 gt521s1 H1:PEM-CS_DUST_LAB3_OPSTATUS: No reply from device within 1000 ms
2025/05/02 07:09:21.140175 gt521s1 H1:PEM-CS_DUST_LAB2_HOLDTIMEMON: No reply from device within 1000 ms
2025/05/02 07:09:22.141296 gt521s1 H1:PEM-CS_DUST_LAB3_HOLDTIMEMON: No reply from device within 1000 ms
 

Images attached to this report