Kyle, Gerardo The following pressures were recorded after opening the gauge isolation valve and waiting for the pressure reading to stabilize (~5 minutes after opening the isolation valve): 11/7/2013 8/20/2003 Dewar 8514371 = 61 microns, 27 microns, 8514372 = 28 microns, 13 microns, 8514374 = 730 microns, 60 microns, 8514375 = 45 microns, 25 microns, 8514376 = 49 microns, 30 microns, 8514377 = 42 microns, 10 microns, 8514379 = 81 microns 26 microns, 8514380 = 58 microns 24 microns The dewar vacuum jackets had been pumped (using a 250L/s turbo pump cart) ~8 months prior to the 8/20/2003 measurements. Note that current LN2 consumption for 8514374 is higher than expected and is likely the result of excessive heat load due to its excessive pressure. Also, this dewar had additional vermiculite? perlite? added to its vacuum jacket since the 8/20/2003 readings.