Beamdumps on HAM1

Betsy, Camilla, Keita

We have started checking beams are sufficiently dumped on the HAM1 beamdumps by checking main beam is easily in the dump and first bounce of the  dumped beam if it's visible.
Progress below. Names as per D1000313 Cartoon.
 
Name Path Beam being dumped Checked?
BD1
POP
LSC POP_A Reflection 
 
BD2
POP
ASC POP_A WFS Reflection 
 
BD3
POP
RM3 Transmission 
 
BD4
POP
BDV2 Reflection 
 
BD5
REFL
BDV1 Reflection
 
BD6a
REFL
M7 Ghost
 
BD6b
REFL
M7 Ghost 
 
BD7
REFL
Unknown, beam from upstream of M2
 
BD8
REFL
M2 Transmission 
YES: photo
BD9
REFL
Unknown, beam from upstream of M2
 
BD10
REFL
ASC REFL_A Reflection
YES: photo
BD11
REFL
ASC REFL_B Reflection 
YES: photo
BD12
REFL
RM2 Transmission 
YES: photo
BD13
REFL
LSC REFL_A Reflection 
YES: photo
BD14
REFL
LSC REFL_B Reflection 
YES: photo
BD15
REFL
M2 Ghost (photo showing distance from main beam)
YES: photo
BD16
REFL
RM1 Transmission 
YES: photo
BD17
PSL
Nozzle Baffle on PSL-HAM1 VP
YES: photo
BD18
PSL
Nozzle Baffle on PSL-HAM1 VP
YES: photo
Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 15:24, Tuesday 13 May 2025 (84376)
Camilla, Betsy, Jenne
All beam dumps are now placed and checked as of this morning 84375, beams were confirmed in each BD apart from BD3 (PM1 transmission) as the beam is too faint to see.
Names as per D1000313 Cartoon, some edits to locations, details of new locations will be in other alog. 
 
Name
Path
Beam being dumped
Checked?
BD1
POP
LSC POP_A Reflection 
YES: photo
BD2
POP
ASC POP_A WFS Reflection, beam clears by ~15mm, photo 
YES: photo
BD3
POP
RM3 Transmission (Beam too weak, estimated placement)
Estimated: photo
BD4
POP
BDV2 Reflection MOVED LOCATIONS
YES: photo
BD5
REFL
BDV1 Reflection, alos catching 2 ghost beams, photo
YES: photo
BD6a
REFL
M14 Ghost MOVED LOCATIONS
YES, see BD6b.
BD6b
REFL
M14 Ghost 
YES: photo
BD7
REFL
Unknown, beam from upstream of M2, MOVED LOCATIONS 
YES: photo
BD8
REFL
M2 Transmission 
YES: photo
BD9
REFL
Unknown, beam from upstream of M2
Not using, dumped by BD8
NewBD
(BD9)
REFL
REFL ASC A low reflection, beam bounces off ISI, photo 
Call this new beamdump BD9 as now not using BD9
YES: photo
BD10
REFL
ASC REFL_A Reflection
YES: photo
BD11
REFL
ASC REFL_B Reflection 
YES: photo
BD12
REFL
RM2 Transmission 
YES: photo
BD13
REFL
LSC REFL_A Reflection 
YES: photo
BD14
REFL
LSC REFL_B Reflection 
YES: photo
BD15
REFL
M2 Ghost (photo showing distance from main beam)
YES: photo
BD16
REFL
RM1 Transmission 
YES: photo
BD17
PSL
Nozzle Baffle on PSL-HAM1 VP
YES: photo
BD18
PSL
Nozzle Baffle on PSL-HAM1 VP
YES: photo
HiPWR
REFL
M14 and M17 Reflection
YES: photo
Images attached to this comment
jennifer.wright@LIGO.ORG - 13:37, Wednesday 14 May 2025 (84392)EPO