Displaying report 1-1 of 1.
Reports until 14:30, Thursday 22 May 2025
H1 SUS (CDS, SUS)
jeffrey.kissel@LIGO.ORG - posted 14:30, Thursday 22 May 2025 (84547)
Greenlighted sush2b HAM1 Software Watchdog (SWWD)
J. Kissel

Now that PM1, RM1, and RM2 are safely functional on the new ISI HAM1, in the h1iopsush2b front-end model, I've
    - set H1:IOP-SUS_${OPTIC}_DK_BYPASS_TIME to 600 seconds H1:IOP-SEI_HAM1_DK_BYPASS_TIME
    - set H1:IOP-SEI_HAM1_DK_BYPASS_TIME to 999999999 (~1e9) seconds, as is seemingly common in most SWWDs
    - hit RESET on each of the SUS DACKILLs and the SEI DACKILL to arm them
    - initialized the PM1 channels in the safe.snap and accepted and monitored their current values
Displaying report 1-1 of 1.