J. Kissel
Now that PM1, RM1, and RM2 are safely functional on the new ISI HAM1, in the h1iopsush2b front-end model, I've
- set H1:IOP-SUS_${OPTIC}_DK_BYPASS_TIME to 600 seconds H1:IOP-SEI_HAM1_DK_BYPASS_TIME
- set H1:IOP-SEI_HAM1_DK_BYPASS_TIME to 999999999 (~1e9) seconds, as is seemingly common in most SWWDs
- hit RESET on each of the SUS DACKILLs and the SEI DACKILL to arm them
- initialized the PM1 channels in the safe.snap and accepted and monitored their current values