PSL Rotation Stage Calibration

After ISCT1 alignment worked wrapped up for the day, I ran a calibration sweep of the PSL rotation stage following the steps outlined in alog79596.

The measurement file, new calibration fit, and screenshot of newly accepted SDFs are attached.

  Power in (W) D B (Minimum power angle) C (Minimum power)
Old Values 94.908 1.990 -24.819 0.000
New Values 95.252 1.990 -25.787

0.000

Images attached to this report
Non-image files attached to this report
Comments related to this report
ryan.short@LIGO.ORG - 17:28, Thursday 22 May 2025 (84555)

Interestingly, even after updating the calibration factors and searching for home on the stage, there are still 2.2W into the IMC with a 2.0W request, so this may need revisiting later.