LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 10:43, Monday 02 June 2025 - last comment - 14:45, Tuesday 03 June 2025(84715)
HAM1 Vent Recovery Update

HAM1 annulus system appears to have issues pumping down, currently its annulus ion pump is railed 10 mA, and the pump cart pumping on this system is sitting at 6.04x10-05 Torr, see attached photo.

Process thus far:

I plan to visit HAM1 to check the rest of the bolts on -Y door and +Y door.

 

Images attached to this report
Comments related to this report
gerardo.moreno@LIGO.ORG - 11:15, Monday 02 June 2025 (84717)VE

HAM1 internal pressure is doing good, see attached photo of the pressure at the SS-500 pump cart pumping on HAM1.  Second attachment is a 5.5 day plot of the pressure internal to HAM1, the large step noted on the plot is due to the introduction of HAM1 ion pump.

Images attached to this comment
gerardo.moreno@LIGO.ORG - 14:45, Tuesday 03 June 2025 (84760)VE

(Jordan V., Gerardo M.)

Today we went and performed the following on the annulus system for HAM1:

  • Disconnected the communications cable, to make sure that the controller is doing its part, used a DB9 connector with the required jumpers, controller behavior did not change.  Removed the bridged connector and reconnected the communication cable. No change noted.
  • Isolated the ion pump via its isolation valve, pressure started to drop, but as soon as the isolation valve was open the pressure goes up.  Valve returned to open, pressure went up.
  • Then we moved to torque some of the bolts on both doors, only the ones that are reachable without bumping on sensitive parts.  Thus far no change on pressure of the annulus.

Now we wait to see the results of today's changes overnight.

 