georgia.mansell@LIGO.ORG - posted 15:42, Monday 02 June 2025 - last comment - 16:18, Monday 02 June 2025(84724)
INP1, CHARD, PRC2 offset tests for input matrix design at 10 W

Julia, Georgia, Sheila, Camilla, Elenna

This might be out of date already as some ASC input matrices have already been updated.

On Sunday we had some locklosses as we powered up from a 0.02Hz oscillation. One known issue is we have taken out some gain out of our CHARD loops. We wanted to check our sensing (input) matrices CHARD, INP1 (which both use the refl wfs) and PRC2 (which uses new in vac pop_x).

We sat at 10W, with the input matrix Sheila found last week:

DOF REFL_A_RF9_I REFL_A_RF45_I REFL_B_RF9_I REFL_B_RF45_I POP_X_I
INP1_P 1   1    
CHARD_P   -1   -0.5  
PRC2_P         -1
INP1_Y 1   1    
CHARD_Y   -1   1  
PRC2_Y         -1

While Elenna measured the phasing using an 8 Hz line, Julia and I added offsets into the loops and looked at the responses in the different sensors.

Some of this is a bit different to what Sheila saw last week, but our ifo alignment has changed. We want to find intrixes that are sensitive to each DOF and ideally are insensitive to the other DOFs.

While writing this another input matrix for INP and CHARD has been put in - https://alog.ligo-wa.caltech.edu/aLOG/index.php?callRep=84711 - and seems to be working so stay tuned.

We plan to work on our gain heirarchy and hopefully power up further soon!

Images attached to this report
Comments related to this report
elenna.capote@LIGO.ORG - 16:18, Monday 02 June 2025 (84726)

These are my radar plots for my 8 Hz dither sensing matrix.

Non-image files attached to this comment