Displaying report 1-1 of 1.
Reports until 17:06, Wednesday 04 June 2025
H1 SQZ
camilla.compton@LIGO.ORG - posted 17:06, Wednesday 04 June 2025 - last comment - 13:42, Thursday 05 June 2025(84794)
SQZ FIS SRC Detuning Dataset

Vicky, Sheila, Kevin, Camilla

Followed plan outlined in 84762 apart from after each SRCL offset change we moved sqz angle roughly close to max sqz to allow SCAN_SQZANG_KHZ enough range to optimize sqz.
After each 3 minutes of data, Kevin and Vicky took a ADF scan.

Increased NLG (needed to adjust SHG waveplate) by taking OPO trans setpoint from 95uW up to 120uW. Measured NLG = 49 (numbers below) following 76542.

Sheila's fitting found that the SRCL  offset (H1:LSC-SRCL1_OFFSET) needed to get a zero degree SRC detuning is -455 counts. Have not accepted in sdf, Sheila added to ISC_LOCK.
Finally we went to FDS and adjusted FC detuning from -34 to -24.

Data attached and saved at /ligo/home/camilla.compton/Documents/sqz/templates/dtt/20250607_SRCdetuning.xml

Type Time (UTC) SRCL Angle DTT Ref
No SQZ 21:37:00 - 21:43:00 N/A N/A ref 0
FIS 22:17:30 - 22:30:30 -191 (CLF-) 242 ref1
FIS 22:35:00 - 22:38:00 -90 (CLF+) 120 ref2
FIS 22:49:30 - 22:52:30 0 (CLF+) 172 ref3
FIS 23:17:00 - 23:20:00 -390 (CLF-) 161 ref4
FIS 23:45:00 - 23:48:00 -455 (CLF-) 143 ref5
FDS ~ 00:00:30 -455 (CLF-) 143 ref6

Note that I left the OPO servo gain at -8, but we have previously left it there (83569) or used -12dB for 120uW OPO trans (83370)

opo_grTrans_ setpoint_uW Amplified Max Amplified Min UnAmp Dark NLG (usual) OPO Gain
120 0.018035 0.000119507 0.000316548 -2e-5 48.9 -8
Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 13:42, Thursday 05 June 2025 (84806)

Here are the plots of this dataset, showing that our SRCL detuning (in gwinc) would be 50 mrad with no digital offset in the SRCL loop.  Kevin tells us that this is a double passed SRCL detuning, so it's really 25 mrad, 1.4 degrees of SRCL detuning. 

Edited to add the last point (and have the correct plots).

Images attached to this comment