Camilla, Ibrahim
Ran A2L before NLN to maxmimize range.
We had to run the a2l script a few times for all QUADs. The first two times, there was no change in gain because the required change vin gain was too much for the script to apply (a safety feature I'm sure). Camilla and I then stepped the gain in the right direction until we were sure that it would be within the adjustability of the script. It took two more runs in order to finally get the range maximized. This improved the range by about 20 MPc. Screenshots attached (the screenshot named initial is before manual stepping).