Reports until 22:36, Thursday 05 June 2025
H1 General (ISC)
oli.patane@LIGO.ORG - posted 22:36, Thursday 05 June 2025 (84845)
Ops Eve Shift End

TITLE: 06/06 Eve Shift: 2330-0500 UTC (1630-2200 PST), all times posted in UTC
STATE of H1: Planned Engineering
INCOMING OPERATOR: None
SHIFT SUMMARY:

The detector is in DOWN for the night and there is a test running right now for Matt for the IMC. The HAM1 cleanroom is on and will be on all night to prep for the +Y door coming off tomorrow morning. Craig and I had turned it on earlier, but then I got worried that it was causing locklosses so I went out and turned it back off, and just now went and turned it back on for the night. It looks like the temperature in that zone (zone 5) was really affected by the cleanroom being turned on (and off) (ndscope1), but the locklosses that we had after it was turned on and back off I don't think should be related to the temperature of the input arm.

After the lockloss earlier in the day during the calibration simulines measurement, we really wanted to get back up and at least check that simulines was running fine and had not caused the lockloss (it didn't look like it did 84836, but still). Unfortunately we weren't able to get back up, as the ifo hasn't really liked me much lately and we had constant locklosses from different places:


LOG:

23:00UTC in NLN_CAL_MEAS
23:14 Started calibration measurement
23:16 Lockloss during calibration measurement, cause unknown
    - Lockloss from ENGAGE_ASC_FOR_FULL_IFO
    - Lockloss from ENGAGE_ASC_FOR_FULL_IFO (from same location)
        - Figured out it was due to the DC6 centering signals not being able to keep up with the movements. Elenna upped the gain for the DC6 loops and this fixed the issue (84843)
    - Lockloss during MOVE_SPOTS due to Craig and me turning on the HAM1 cleanroom
    - Lockloss from MOVE_SPOTS due to a PRC1/2 P ringup at 0.27 Hz
    - Lockloss from ENGAGE_ASC_FOR_FULL_IFO due to 84843
    - I turned the HAM1 cleanroom back off in case it was affecting our ability to lock
    - Lockloss from TRANSITION_FROM_ETMX due to 1 Hz oscillation in DHARD P and PRC2 P from ringup (84844)

Start Time System Name Location Lazer_Haz Task Time End
00:11 PEM Robert LVEA n Listening for noise 00:24
01:16   Oli, Craig LVEA n Turning on HAM1 cleanroom 01:23
02:13   Oli LVEA n Turning HAM1 cleanroom off 02:19
05:12   Oli LVEA n Turning HAM1 cleanroom on 05:16
Images attached to this report