Laser Status:
NPRO output power is 1.85W
AMP1 output power is 70.39W
AMP2 output power is 140.3W
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 18 days, 0 hr 26 minutes
Reflected power = 23.16W
Transmitted power = 105.5W
PowerSum = 128.6W
FSS:
It has been locked for 0 days 1 hr and 12 min
TPD[V] = 0.801V
ISS:
The diffracted power is around 3.8%
Last saturation event was 0 days 13 hours and 21 minutes ago
Possible Issues:
PMC reflected power is high