Reports until 12:44, Friday 06 June 2025
H1 CDS (ISC)
filiberto.clara@LIGO.ORG - posted 12:44, Friday 06 June 2025 (84867)
ISCT1 and IOT2L Cameras and Table Enclosure Lights

WP 12572
The 24V power for the table enclosure lights on ISCT1 and IOT2L were rerouted through the Interlock/Lighting patch panel D1201535. Panel has a power switch and Status On/Off indicator. The dimmer functional was left in place and requires both the panel power switch and dimmer in the ON position for LEDs to power on.

WP 12574
New GigE camera was installed in ISCT1, alog 84866. Camera is connected to the new CER switch, port 12. Three ports on the SUS-R1 patch panel are available for new GigE cameras.

Power to the three IOT2L analog cameras was disconnected:

  1. WFSA
  2. WFSB
  3. REFL

Power cable did not have a dedicated Feedthrough. Analog cameras are no longer used. The 18V power cable is on top of IOT2L if power needs to be reconnected.