The PID controller for EY CP7 cryopump has opened its liquid level control valve (LLCV) to 100% but this is proving to be insufficient to maintain a 92% fill level in the pump. The pump level started deviating from nominal around 4pm Friday 06jun2025 PDT.
Vacuum team is investigating.
Note: no alarms issued at this point, the cyropump alarm level is < 80%, CP7 currently is at 85%.
I had noted the LLCV issue for CP7 here. Tomorrow we'll attempt to re-zero the LLCV actuator.