Some Observe.snap SDF clearing

I'm working on going through some Observe SDFs, so that we're ready for observing soon.

Jim is currently working on going through many of the SEI SDFs.  The rest of the diffs I need to check with other commissioners to be sure about before we clear them, but I think we're getting close to having our SDFs cleared!

Images attached to this report
Comments related to this report
camilla.compton@LIGO.ORG - 14:05, Tuesday 10 June 2025 (84935)

h1tcshws sdfs attached.

Reverted BaffePDs to what they were 3 months ago, attached, unsure why they would have changed.

SQZ ADF frequency sdfs accepted, we do not know why these would have been accepted at the values -600 that they've been at some of the past 2 weeks.

Images attached to this comment
elenna.capote@LIGO.ORG - 13:35, Tuesday 10 June 2025 (84936)

ASC SDFs were from changes to DC6, cleared.

elenna.capote@LIGO.ORG - 13:39, Tuesday 10 June 2025 (84937)

Cleared these SDFs for the phase changes for LSC REFL A and B.

Images attached to this comment
sheila.dwyer@LIGO.ORG - 16:14, Tuesday 10 June 2025 (84945)

I trended these, and see that FM2 was on in all three of these last time we were in observing, so these must have been erroneously accepted in the observing snap.

Images attached to this comment
camilla.compton@LIGO.ORG - 08:23, Wednesday 11 June 2025 (84965)

I also accepted the HAM7_DK_BYPASS time from 1200 to 999999 after checking with Dave, as attached.