H1 dropped observing from 18:30 to 19:00 UTC for regularly scheduled calibration measurements, which ran without issue. A screenshot of the calibration monitor medm and the calibration report are attached.
Broadband runtime: 18:30:45 to 11:35:54 UTC
Simulines runtime: 18:36:41 to 18:59:56 UTC
We had to rerun the report to account for pro-spring in the model. Calibration looks better now -- sensing model is within 2% above 20 Hz and 5% below 20 Hz, report attached. I also updated the .ini file to now account for the pro-spring behavior.
More detailed steps:
is_pro_spring
to True
at the pydarm_H1.ini
in report 20250614T183642Z
20250614T183642Z
(in terminal, ran $pydarm report --regen --skip-gds 20250614T183642Z
)pydarm_H1.ini
file at /ligo/groups/cal/H1/ifo
as pydarm_H1.ini.250610
to save previous configuration./ligo/groups/cal/H1/ifo/pydarm_H1.ini
, set is_pro_spring
to True