ryan.short@LIGO.ORG - posted 12:06, Saturday 14 June 2025 - last comment - 09:47, Monday 16 June 2025(85045)
Broadband and Simulines Calibration Sweeps

H1 dropped observing from 18:30 to 19:00 UTC for regularly scheduled calibration measurements, which ran without issue. A screenshot of the calibration monitor medm and the calibration report are attached.

Broadband runtime: 18:30:45 to 11:35:54 UTC

Simulines runtime: 18:36:41 to 18:59:56 UTC

Images attached to this report
Non-image files attached to this report
Comments related to this report
francisco.llamas@LIGO.ORG - 09:47, Monday 16 June 2025 (85073)

We had to rerun the report to account for pro-spring in the model. Calibration looks better now -- sensing model is within 2% above 20 Hz and 5% below 20 Hz, report attached. I also updated the .ini file to now account for the pro-spring behavior.

More detailed steps:

  1. Set is_pro_spring to True at the pydarm_H1.ini in report 20250614T183642Z
  2. Rerun report 20250614T183642Z (in terminal, ran $pydarm report --regen --skip-gds 20250614T183642Z)
  3. Copied current pydarm_H1.ini file at /ligo/groups/cal/H1/ifo as pydarm_H1.ini.250610 to save previous configuration.
  4. In /ligo/groups/cal/H1/ifo/pydarm_H1.ini , set is_pro_spring to True

 