Displaying report 1-1 of 1.
Reports until 09:06, Thursday 19 June 2025
H1 CAL
ibrahim.abouelfettouh@LIGO.ORG - posted 09:06, Thursday 19 June 2025 - last comment - 15:28, Friday 20 June 2025(85179)
Calibration Sweep 06/19 - Failed - Lockloss during Simulines -

Headless Start: 1434382652

2025-06-19 08:42:24,391 bb measurement complete.
2025-06-19 08:42:24,391 bb output: /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20250619T153714Z.xml
2025-06-19 08:42:24,392 all measurements complete.
Headless End: 1434382962

Simulines Start: 1434383042

2025-06-19 15:55:14,510 | INFO | Drive, on L3_SUSETMX_iEXC2DARMTF, at frequency: 8.99, and amplitude 0.53965, is finished. GPS start and end time stamps: 1434383704, 1434383727
2025-06-19 15:55:14,510 | INFO | Scanning frequency 10.11 in Scan : L3_SUSETMX_iEXC2DARMTF on PID: 2067315
2025-06-19 15:55:14,511 | INFO | Drive, on L3_SUSETMX_iEXC2DARMTF, at frequency: 10.11, is now running for 28 seconds.
2025-06-19 15:55:17,845 | ERROR | IFO not in Low Noise state, Sending Interrupts to excitations and main thread.
2025-06-19 15:55:17,846 | ERROR | Ramping Down Excitation on channel H1:LSC-DARM1_EXC
2025-06-19 15:55:17,846 | ERROR | Ramping Down Excitation on channel H1:SUS-ETMX_L2_CAL_EXC
2025-06-19 15:55:17,846 | ERROR | Ramping Down Excitation on channel H1:SUS-ETMX_L3_CAL_EXC
2025-06-19 15:55:17,846 | ERROR | Ramping Down Excitation on channel H1:CAL-PCALY_SWEPT_SINE_EXC
2025-06-19 15:55:17,846 | ERROR | Ramping Down Excitation on channel H1:SUS-ETMX_L1_CAL_EXC
2025-06-19 15:55:17,846 | ERROR | Aborting main thread and Data recording, if any. Cleaning up temporary file structure.
PDT: 2025-06-19 08:55:22.252424 PDT
UTC: 2025-06-19 15:55:22.252424 UTC
Simulines End GPS: 1434383740.252424

Could not generate a report using the wiki instructions (probably due to incomplete sweep).

Images attached to this report
Comments related to this report
elenna.capote@LIGO.ORG - 15:28, Friday 20 June 2025 (85209)

Here is a screenshot of the broadband measurement. The calibration still looks good!

Images attached to this comment