Displaying report 1-1 of 1.
Reports until 10:08, Monday 23 June 2025
H1 SQZ
camilla.compton@LIGO.ORG - posted 10:08, Monday 23 June 2025 - last comment - 13:19, Monday 23 June 2025(85237)
SQZ FIS Data set
Sheila, Camilla

After Kevin did some 10kHz SQZ ADF scans, we look some SQZ FIS data. The best ASQZ and SQZ didn't look as good as usual, but we found our NLG was low. We didn't check the NLG first so that we had a direct comparison to Kevin's dataset.

Roughly following 83594 with slightly different data taken.

DTT saved as camilla.compton/Documents/sqz/templates/dtt/20250623_FIS.xml and screenshot attached.

Type Time (UTC) Angle DTT Ref
FIS Mean SQZ 15:53:00 - 15:56:00 N/A ref 1
FIS + Mid SQZ (aimed +7dB @1kHz) 15:58:30 - 16:01:30 (-) 242 ref 2
FIS - Mid SQZ (aimed +7dB @1kHz) 16:03:00 - 16:06:00 (-) 94 ref 3
FIS ASQZ 16:08:30 - 16:11:30 (-) 63 ref 4
FIS SQZ 16:13:30 -16:16:30 (-) 122 ref 5
No SQZ 16:26:00 -  16:46:00 N/A ref 0
 
Checked while in NO SQZ, also adjusted the SHG pump waveplates to reduce the HAM7 rejected power, but I did this less than a week ago:
OPO Setpoint Amplified Max Amplified Min UnAmp Dark NLG OPO Gain Note
95uW 0.001504 6.71 e-5 0.0002828 -2.52e-5 4.9 -8 Without changing OPO temp
95uW 0.002317 8.69e-5 0.0002828 -2.52e-5 7.5 -8 With changing OPO temp

Last week 85000, NLG was measured to be 18.9, now it's measured to be <8. We will look into why is it low.

Ran SCAN_SQZANG_FDS before doing a CO2 step.

Images attached to this report
Comments related to this report
sheila.dwyer@LIGO.ORG - 13:19, Monday 23 June 2025 (85252)

 I measured the NLG with a few different opo pump powers to follow up on Camilla's observation that the NLG was low. 

OPO trans power [uW] maximum minimum nlg
95 2.26e-3 8.195e-5 8.5
80 2.25e-3 8.795e-5 8.2
60 1.902e-3 8.866e-5 6.93

dark level 2e-5, unamplified 2.914e-4

I tried to lower the seed power to see if we have some pump depletion making us underestimate the NLG, but the half wave plate was already set to minimize the seed power.

Since the NLG looks to be about the same with 80 uW as 95uW, I set it to 80 uW for now. 

Displaying report 1-1 of 1.