I have installed the level 2.0 HSTS damping filters into the FC1 damping bank. I also accepted the changes in SDF.
Open loop gain TFs have been taken of all FC1 dofs with the new damping design installed and are in the SusSVN.
Here're the templates associated with the open loop gain TFs of FC1's M1 damping loops that Elenna mentions:
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/FC1/SAGM1/Data/
2022-10-14_2330_H1SUSFC1_M1_CDBIOState_1_WhiteNoise_L_0p01to100Hz_OpenLoopGainTF_new.xml
2022-10-14_2330_H1SUSFC1_M1_CDBIOState_1_WhiteNoise_P_0p01to100Hz_OpenLoopGainTF_new.xml
2022-10-14_2330_H1SUSFC1_M1_CDBIOState_1_WhiteNoise_R_0p01to100Hz_OpenLoopGainTF_new.xml
2022-10-14_2330_H1SUSFC1_M1_CDBIOState_1_WhiteNoise_T_0p01to100Hz_OpenLoopGainTF_new.xml
2022-10-14_2330_H1SUSFC1_M1_CDBIOState_1_WhiteNoise_V_0p01to100Hz_OpenLoopGainTF_new.xml
2022-10-14_2330_H1SUSFC1_M1_CDBIOState_1_WhiteNoise_Y_0p01to100Hz_OpenLoopGainTF_new.xml