Displaying report 1-1 of 1.
Reports until 02:14, Wednesday 25 June 2025
LHO VE (VE)
gerardo.moreno@LIGO.ORG - posted 02:14, Wednesday 25 June 2025 (85317)
Annulus Ion Pump for BSC1 Rails

Signal railed about 5:18 PM local time, I checked trend data for PT120 and PT180 and no pressure rise noted inside the main volume.  Attached is 3 day trend of the pump behavior, very glitchy for a long while already.

System will be evaluated as soon as possible.  AIP last replaced on 2015, see aLOG 18261.

Images attached to this report
Displaying report 1-1 of 1.