OMC OL values

I logged the OMC OLs for Jeff B:

prettyOSEMgains('H1','OMC')

M1T1 20997 1.429 -10499
M1T2 21195 1.415 -10598
M1T3 23620 1.270 -11810
M1LF 22733 1.320 -11366
M1RT 21724 1.381 -10862
M1SD 22193 1.352 -11096
 

Comments related to this report
mark.barton@LIGO.ORG - 11:05, Thursday 14 November 2013 (8537)

I entered the OSEMINF offset and gain values, entered the correct magnet signs in COILOUTF and did a new safe.snap. I had to do this manually because the Matlab script doesn't cope with the fact that the autoBurt.req has three bad lines due to a bug in RTCDS 2.8 leading to the folllowing errors:

H1:SUS-OMC_DACKILL_PANIC.HSV ... not connected so no ca_array_get_callback()
H1:SUS-OMC_DACKILL_PANIC.LSV ... not connected so no ca_array_get_callback()
H1:SUS-OMC_DACKILL_PANIC.LOW ... not connected so no ca_array_get_callback()
 