Displaying report 1-1 of 1.
Reports until 12:06, Thursday 14 November 2013
H1 SUS
arnaud.pele@LIGO.ORG - posted 12:06, Thursday 14 November 2013 (8542)
ETMX TF results

ETMX transfer functions Phase 3a after 2nd round of alignment yesterday (cf alog 8503 and 8525) have succesfully been completed, and they show an excellent agreement with the previous measurements taken after the first alignment.

Results attached are described below

(1) M0-M0 undamped TF comparison with the model

(2) R0-R0 undamped TF comparison with the model

(3) M0-M0 undamped TF comparison with the first phase 3a measurement, and the model

(4) R0-R0 undamped TF comparison with the first phase 3a measurement, and the model

Data results and scripts have been commited to the svn

Non-image files attached to this report
Displaying report 1-1 of 1.