filiberto.clara@LIGO.ORG - posted 12:33, Tuesday 01 July 2025 - last comment - 15:58, Wednesday 02 July 2025(85463)
Top Sat Amps Modified for PRM/PR3/SR3/SRM/BS

WP 12640
ECR E2400330
Drawing D0901284-v5
T2500232

Today we started upgrading the SUS Sat Amps per ECR E2400330. Modification improves the whitening stage to reduce ADC noise from 0.05 to 10 Hz. The following units were replaced with modified units:

Suspension Old New
PRM TOP S1100067 S1100168
PRM/PR3 S1000275 S1100173
PR3 S1100175 S1100183
     
BS TOP S1100090 S1000291
BS RT/SD  S1100155 S1100066
     
SR3 Top S1000296 S1000284
SR3/SRM S1100178 S1000290
SRM S1100136 S1100068

F. Clara, C. Gray,  J. Kissel, O. Patane, M.Pirello

Comments related to this report
jeffrey.kissel@LIGO.ORG - 13:48, Wednesday 02 July 2025 (85504)CDS
Here's the characterization data and fit results for S1100168, assigned to PRM M1's T1T2T3LF OSEMs (what Fil refers to as just "PRM TOP" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
        plotresponse_S1100168_PRM_M1_T1T2T3LF_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
PRM      M1       S1100168         CH1                T1           0.0969 : 5.30   120.10        zpk([5.3],[0.0969],1,"n")
                                   CH2                T2           0.0950 : 5.20   120.25        zpk([5.2],[0.095],1,"n")
                                   CH3                T3           0.0950 : 5.18   120.25        zpk([5.18],[0.095],1,"n")
                                   CH4                LF           0.0940 : 5.15   120.00        zpk([5.15],[0.094],1,"n")

The attached plot and machine readable .txt file are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

Even though I've fit for the transimpedance gain, I've elected *not* to include a gain in the foton design string relative to "ideal," as there are more scale factors in play that determine the overall [(meters)/(ADC cts)] scale in the calibration of the OSEMs (LED light power, PD response, any cable loss, ADC channel gain, etc.). Determining this overall scale is better left to different methods, a la LHO:84548, which we (eventually) anticipate doing for all SUS with ECR E2400330 upgraded satamps.

Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 14:05, Wednesday 02 July 2025 (85505)CDS
Here's the characterization data and fit results for  S1100173 , assigned to PRM/PR3 M1's RTSD/T1T2 OSEMs (what Fil refers to as just "PRM/PR3" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
        plotresponse_S1100173_PRMPR3_M1_RTSDT1T2_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
PRMPR3   M1       S1100173         CH1                RT           0.0969 : 5.3    120.00        zpk([5.3],[0.0969],1,"n")
                                   CH2                SD           0.0955 : 5.22   120.00        zpk([5.22],[0.0955],1,"n")
                                   CH3                T1           0.0975 : 5.35   119.75        zpk([5.35],[0.0975],1,"n")
                                   CH4                T2           0.0975 : 5.33   120.25        zpk([5.33],[0.0975],1,"n")

The attached plot and machine readable .txt file version of the above table are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

See above LHO:85504 comment as to why an account of the fit transimpedance was not included in the foton design string gain.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 14:06, Wednesday 02 July 2025 (85506)CDS
Here's the characterization data and fit results for  S1100183 , assigned to PR3 M1's T3LFRTSD OSEMs (what Fil refers to as just "PR3" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
        plotresponse_S1100183_PR3_M1_T3LFRTSD_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
PR3      M1       S1100183         CH1                T3           0.0950 : 5.19   120           zpk([5.19],[0.095],1,"n")
                                   CH2                LF           0.0945 : 5.17   120           zpk([5.17],[0.0945],1,"n")
                                   CH3                RT           0.0940 : 5.14   120           zpk([5.14],[0.094],1,"n")
                                   CH4                SD           0.0955 : 5.24   120           zpk([5.24],[0.0955],1,"n")

The attached plot and machine readable .txt file version of the above table are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

See above LHO:85504 comment as to why an account of the fit transimpedance was not included in the foton design string gain.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 14:11, Wednesday 02 July 2025 (85507)
Here's the characterization data and fit results for  S1000291 , assigned to BS M1's F1F2F3LF OSEMs (what Fil refers to as just "BS TOP" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
        plotresponse_S1000291_BS_M1_F1F2F3LF_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
BS       M1       S1000291         CH1                F1           0.096 : 5.19    121.50        zpk([5.19],[0.096],1,"n")
                                   CH2                F2           0.096 : 5.23    120.75        zpk([5.23],[0.096],1,"n")
                                   CH3                F3           0.096 : 5.26    120.00        zpk([5.26],[0.096],1,"n")
                                   CH4                LF           0.095 : 5.2     120.00        zpk([5.2],[0.095],1,"n")

The attached plot and machine readable .txt file version of the above table are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

See above LHO:85504 comment as to why an account of the fit transimpedance was not included in the foton design string gain.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 14:18, Wednesday 02 July 2025 (85508)
Here's the characterization data and fit results for  S1100066 , assigned to BS M1's RTSDxxxx OSEMs (with the 2x "xx" representing the unused channels not connected to anything in-vacuum. Fil refers to this as just "BS RT/SD" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
        plotresponse_S1100066_BS_M1_RTSDxxx_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
BS       M1       S1100066         CH1                RT           0.0960 : 5.19   121.50        zpk([5.19],[0.096],1,"n")
                                   CH2                SD           0.0940 : 5.11   120.75        zpk([5.11],[0.094],1,"n")
                                   CH3                xx           0.0935 : 5.12   120.25        zpk([5.12],[0.0935],1,"n")
                                   CH4                xx           0.0965 : 5.26   120.50        zpk([5.26],[0.0965],1,"n")

The attached plot and machine readable .txt file version of the above table are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

See above LHO:85504 comment as to why an account of the fit transimpedance was not included in the foton design string gain.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 14:22, Wednesday 02 July 2025 (85509)
Here's the characterization data and fit results for  S1000284 , assigned to SR3 M1's T1T2T3LF OSEMs (Fil refers to this as just "SR3 Top" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
        plotresponse_S1000284_SR3_M1_T1T2T3LF_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
SR3      M1       S1000284         CH1                T1           0.0960 : 5.25   120.1         zpk([5.25],[0.096],1,"n")
                                   CH2                T2           0.0950 : 5.2    120.1         zpk([5.2],[0.095],1,"n")
                                   CH3                T3           0.0960 : 5.27   120.1         zpk([5.27],[0.096],1,"n")
                                   CH4                LF           0.0945 : 5.16   120.5         zpk([5.16],[0.0945],1,"n")

The attached plot and machine readable .txt file version of the above table are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

See above LHO:85504 comment as to why an account of the fit transimpedance was not included in the foton design string gain.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 14:30, Wednesday 02 July 2025 (85511)
Here's the characterization data and fit results for  S1000290 , assigned to SR3/SRM M1's RTSD/T1T2 OSEMs (Fil refers to this as just "SR3/SRM" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
        plotresponse_S1000290_SR3SRM_M1_RTSDT1T2_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
SR3/SRM  M1       S1000290         CH1                RD           0.0955 : 5.21   120.1         zpk([5.21],[0.0955],1,"n")
                                   CH2                SD           0.0940 : 5.13   120.5         zpk([5.13],[0.094],1,"n")
                                   CH3                T1           0.0930 : 5.07   120.5         zpk([5.07],[0.093],1,"n")
                                   CH4                T2           0.0950 : 5.2    120.1         zpk([5.2],[0.095],1,"n")

The attached plot and machine readable .txt file version of the above table are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

See above LHO:85504 comment as to why an account of the fit transimpedance was not included in the foton design string gain.
Non-image files attached to this comment
jeffrey.kissel@LIGO.ORG - 15:58, Wednesday 02 July 2025 (85513)CDS
Here's the characterization data and fit results for  S1100068 , assigned to SRM M1's T3LFRTSD OSEMs (Fil refers to this as just "SRM" above).
The data was taken per methods described in T080062-v3.
The data was processed and fit using 
    ${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/
         plotresponse_S1100068_SRM_M1_T3LFRTSD_20250630.m

Explicitly, the fit to the whitening stage zero and pole, the transimpedance feedback resistor, and foton design string are
Optic    Stage    Serial_Number    Channel_Number     OSEM_Name    Zero_Pole_Hz    R_TIA_kOhm    Foton_Design
SRM      M1       S1100068         CH1                T3           0.094 : 5.14    120.10        zpk([5.14],[0.094],1,"n")
                                   CH2                LF           0.095 : 5.18    120.25        zpk([5.18],[0.095],1,"n")
                                   CH3                RT           0.096 : 5.23    120.75        zpk([5.23],[0.096],1,"n")
                                   CH4                SD           0.096 : 5.29    119.00        zpk([5.29],[0.096],1,"n")

The attached plot and machine readable .txt file version of the above table are also found in
${SusSVN}/trunk/electronicstesting/lho_electronics_testing/satamp/ECR_E2400330/Scripts/

See above LHO:85504 comment as to why an account of the fit transimpedance was not included in the foton design string gain.
Non-image files attached to this comment