IAS: D. Cook, J. Oberling
SUS: B. Weaver
Fine aligned the ITMx pitch this afternoon.
-
Pitch: 976 µrad down
-
Target: 971 µrad down (619 µrad optic pitch + 352 µrad FC correction)
-
Error: 5 µrad down
-
Tolerance: ±50 µrad
SUS will run transfer functions to confirm the absence of rubbing.
We also aligned the reaction chain - it now sits within the pitch tolerance at ~1.1mRad up, but we expect it to point to ~600mRad down after we take the heavy FC sheet off of the CP-AR surface.
Also to note - I closed the Ring Heater around the ITMx.