Reports until 07:39, Thursday 03 July 2025
LHO General
corey.gray@LIGO.ORG - posted 07:39, Thursday 03 July 2025 - last comment - 08:49, Thursday 03 July 2025(85518)
Thurs DAY Ops Transition

TITLE: 07/03 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 2mph 3min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.07 μm/s
QUICK SUMMARY:

H1 had (3)  ~2-hr locks overnight (looks like no wake-up calls for Oli!).  H1 is currently locking (just powered up to 25W).  Winds stepped down at about 130amPDT (837utc).

Today is Thurs Commissioning from 8am-12pmPDT (15-19utc).  H1 won't be thermalized until about 1030-11amPDT, so no calibration...hopefully H1 short locks won't be an issue for calibration where we need H1 thermalized 3hrs.  Robert and Sheila will probably start their work

Comments related to this report
elenna.capote@LIGO.ORG - 08:23, Thursday 03 July 2025 (85520)

Just want to add a note that all three locks from last night were nearly the exact same length, 1:51, and the locklosses are all tagged with "PI monitor". A ring heater was changed yesterday, 85514, which may have caused this problem.

corey.gray@LIGO.ORG - 08:29, Thursday 03 July 2025 (85521)SUS, TCS

Additionally, each of the 3 locklosses were from PI28/29 according to VerbalAlarms (the Ring Heater change yesterday was made to address PI24 which riddled our 7hr lock for the last hour before the ultimate lockloss).

elenna.capote@LIGO.ORG - 08:49, Thursday 03 July 2025 (85522)

While we have been running PIMON live on nuc25, it appears the data from the lockloss hasn't been saved. The newest file in the /ligo/gitcommon/labutils/pimon/locklosses folder is from December 2024. I'm not sure what's going on. We think our problem is an 80 kHz PI, so this would be useful data to have.