anthony.sanchez@LIGO.ORG - posted 07:47, Friday 04 July 2025 - last comment - 09:17, Friday 04 July 2025(85545)
Friday July 4th Ops day shift start

TITLE: 07/04 Day Shift: 1430-2330 UTC (0730-1630 PST), all times posted in UTC
STATE of H1: Lock Acquisition
OUTGOING OPERATOR: Oli
CURRENT ENVIRONMENT:
    SEI_ENV state: CALM
    Wind: 5mph Gusts, 2mph 3min avg
    Primary useism: 0.01 μm/s
    Secondary useism: 0.06 μm/s
QUICK SUMMARY:
When I Arrived H1 was in Nominal_LOW NOISE, but not Observing. Range: 120
SQZ_LO_LR was stuck in Locking_OMC[15] with an error message: Beam diverter is open but no 3 MHz on OMC, could be an alignment problem.

Things i have done already.
Init SQZ_LO_LR & then SQZ MANAGER

 

 

Comments related to this report
anthony.sanchez@LIGO.ORG - 09:17, Friday 04 July 2025 (85546)

"Beam diverter is open but no 3MHz on OMC, could be an alignment problem" - trend ZM4,5,6 alignment while trending H1:SQZ-OMC_TRANS_RF3_DEMOD_RFMON. If the alignment of any has changed, move them back slowly while aiming to increase H1:SQZ-OMC_TRANS_RF3_DEMOD_RFMON.

Found here:Troubleshooting SQZr from Ops wiki.
Slider values had not changed at all over the course of several hours but the osems had. 
Sheila called while I was setting up my ndscope to check the Osems and had mentioned that ZM5 likely didnt matter as it is ZM4 and 6 that get moved.

She had me clear the history on H1:SUS-ZM{4 & 6}_M1_LOCK_{L P Y}_OUTPUT before I  adjusted the ZM4 & 6 sliders.
H1:SUS-ZM4_M1_LOCK_L_GAIN were changed after turning up the TRamps for H1:SUS-ZM{4 & 6}_M1_LOCK_{L, P, Y}_GAIN
Then I cleared the History on those locking filters.

Once that was done I simply just move them back slowly while aiming to increase H1:SQZ-OMC_TRANS_RF3_DEMOD_RFMON as the Trouble shooting doc suggested.

Then as Sheila advised I ren SQZ_MAN through Scan_Alignment_FDS[105] wich further touched up alignment. 
After accepting SDF diffs.

We got back to Observing at 15:57 UTC @ 149 Mpc.
 

Images attached to this comment