Displaying report 1-1 of 1.
Reports until 23:03, Saturday 05 July 2025
H1 SQZ
ryan.crouch@LIGO.ORG - posted 23:03, Saturday 05 July 2025 (85568)
OPS OWL assistance

I got a call 05:39 UTC and found the SQZ to be struggling to lock, notifying "LO RAILING", I consulted the alog and found a report alog81027 from Vicky and myself where we faced the same issue. I checked the FC alignment and found it to be the same, I then checked the ZMs and found ZM6 to be in a very different spot, especially in pitch. I brought back ZM6s alignment which let the LO lock but our range was terrible, I then adjusted the OPO temperature and ran a SQZ ang adjust as it was reporting that the angle was out of range then everything looked good.

05:55 UTC Observing

Displaying report 1-1 of 1.