Reports until 17:31, Thursday 14 November 2013
H1 SUS (SUS)
rich.abbott@LIGO.ORG - posted 17:31, Thursday 14 November 2013 (8561)
ESD
Looked at the connections to the H1 ITMX ESD starting at the airside of the vacuum feedthrough.  There is no evidence that there is any connection being made to ANYTHING inside the vacuum system.  In the process of analyzing this, I installed the Delrin insert on the airside of the vacuum feedthrough and very carefully connected the coaxial cables.  Truly no way it's connected on the inside of the vacuum flange.