oli.patane@LIGO.ORG - posted 11:30, Tuesday 08 July 2025 (85618)
SUS OLTFs taken for ETMY M0, SR2, FC1, FC2

I've taken open loop transfer functions for the rest of the suspensions that we didn't have any (or had very old) for: ETMY M0, SR2, FC1, and FC2

These were all taken before the sat amp swap (0.4:10 satamp)

ETMY M0
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
    - optic align offsets off, L2->R0 damping off, etc
*Templates copied from ETMX M0
- V excitation amplitude had to be reduced from 5 to 1
- Y excitation amplitude had to be reduced from 25 to 5
/ligo/svncommon/SusSVN/sus/trunk/QUAD/H1/ETMY/SAGM0/Data/2025-07-08_1630_H1SUSETMY_M0_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGainTF.xml r12399

SR2
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
    - optic align offsets off
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/SR2/SAGM1/Data/2025-07-08_1645_H1SUSSR2_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGainTF.xml r12401

FC1
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
    - optic align offsets off
    - *Measurements taken with HAM7 tripped because it was tripping all morning and I didn't want it to trip during measurements
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/FC1/SAGM1/Data/2025-07-08_1730_H1SUSFC1_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGain.xml r12402

FC2
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
    - optic align offsets off
*Templates were copied over from FC1
- R excitation amplitude had to be reduced from 20 to 10
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/FC2/SAGM1/Data/2025-07-08_1745_H1SUSFC2_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGain.xml r12404