I've taken open loop transfer functions for the rest of the suspensions that we didn't have any (or had very old) for: ETMY M0, SR2, FC1, and FC2
These were all taken before the sat amp swap (0.4:10 satamp)
ETMY M0
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
- optic align offsets off, L2->R0 damping off, etc
*Templates copied from ETMX M0
- V excitation amplitude had to be reduced from 5 to 1
- Y excitation amplitude had to be reduced from 25 to 5
/ligo/svncommon/SusSVN/sus/trunk/QUAD/H1/ETMY/SAGM0/Data/2025-07-08_1630_H1SUSETMY_M0_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGainTF.xml r12399
SR2
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
- optic align offsets off
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/SR2/SAGM1/Data/2025-07-08_1645_H1SUSSR2_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGainTF.xml r12401
FC1
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
- optic align offsets off
- *Measurements taken with HAM7 tripped because it was tripping all morning and I didn't want it to trip during measurements
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/FC1/SAGM1/Data/2025-07-08_1730_H1SUSFC1_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGain.xml r12402
FC2
- Measurements taken with suspension in HEALTH_CHECK but with damping loops on
- optic align offsets off
*Templates were copied over from FC1
- R excitation amplitude had to be reduced from 20 to 10
/ligo/svncommon/SusSVN/sus/trunk/HSTS/H1/FC2/SAGM1/Data/2025-07-08_1745_H1SUSFC2_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGain.xml r12404