Jennie, Rahul
On Tuesday Rahul and I took the measurements for the horizontal coupling in the ISS array currently on the optical table.
The QPD read 9500 e-7 W.
The X position was 5.26 V, the Y position was -4.98 V.
PD | DC Voltage [mV] pk-pk | AC Voltage [mV] pk-pk |
1 | 600 | 420 |
2 | 600 | 380 |
3 | 600 | 380 |
4 | 600 | 420 |
5 | 800 | 540 |
6 | 800 | 500 |
7 | 600 | 540 |
8 | 800 | 540 |
After thinking about this data I realise we need to retake it as we should record the mean value for the DC coupled measurements. This was with a 78V signal applied from the PZT driver and an input dither signal of 2 Vpp at 100Hz on the oscilloscope and I think 150 mA pump current on the laser.
Rahul, Jennie W
Yesterday we went back into the lab and retook the DC and AC measurements while horizontal dither was on while measuring using the 'mean' setting and without changing the overall input pointing from what it was in the above measurement.
QPD sum is roughly 5V.
Next time we need to plug in the second axis of the PZT driver so as to take the dither coupling measurement in the vertical direction.