Reports until 13:07, Monday 21 July 2025
H1 ISC
elenna.capote@LIGO.ORG - posted 13:07, Monday 21 July 2025 (85894)
Calibration lines in DHARD

The calibration lines still appear in DHARD P and Y to varying degrees. Here is a higher resolution spectrum of DHARD P and Y, with OMC DCPD sum for reference. DHARD P shows all four calibration lines (L1 15.6 Hz, L2 16.4 Hz, L3 17.6 Hz and PCAL 17.1 Hz) between 15-18 Hz, while DHARD Y only seems to show the 16.4 Hz line in L2.

Today, Sheila and I also looked at the phasing of AS A/B 45 WFS. Our thinking was that a DARM line could be phased to be mainly in Q for each segment. The segments all show the calibration lines prominently, so they can be used as a phasing reference. It was immediately apparently that for all segments of both WFS, the DARM line is at least a factor of 2 higher in I than in Q. We decided not to make any phasing changes at this time, since rephasing these lines into Q would change the sign dramatically and may impact locking when we put DARM on RF.

Images attached to this report