Displaying report 1-1 of 1.
Reports until 12:33, Tuesday 22 July 2025
H1 AOS
camilla.compton@LIGO.ORG - posted 12:33, Tuesday 22 July 2025 - last comment - 13:11, Friday 22 August 2025(85917)
SQZT7 Beam Profiling with different ZM4 and ZM5 PSAMS Settings

Leo, Jennie, Camilla, WP 12694

Jennie followed instructions for set up in 85775. We removed the SQZ beam iris at the bottom of the LPM (added for alignment capture during OFI vent work). Then took beam profiles in this SQZ path of the SEED beam with various PSAMS settings, adjusted PSAMS settings as in 85775 and used the servo for nominal settings. All data is attached. Jennie then reverted settings back to nominal.

Non-image files attached to this report
Comments related to this report
leendert.schrader@LIGO.ORG - 12:01, Thursday 14 August 2025 (86365)
Leo, Jennie W., Camilla

The attached pdf contains all the beam q parameters fitted to the collected beam width data. Only the 13.5% data was fitted, as the D4S data was too inconsistent to obtain confident q values.

Fitting was performed with the a la mode beamPath.fitBeamWidth function. 
The attached q parameters were individually plotted using a la mode and verified for their data-fitting accuracy.

As mentioned in the document, all q parameters are located immediately after the interaction with ZM5 (through the view of BM4 -> ZM4 -> ZM5 beam travel).
Non-image files attached to this comment
leendert.schrader@LIGO.ORG - 13:11, Friday 22 August 2025 (86519)
Leo, Jennie W., Camilla

Attached is a plot of the q manifold from the q parameter data, which allows for characterizing the beam smoothly with respect to ZM4/5 strain gauge voltage values.
The image is taken from the presentation uploaded to T2500228. The real plot will likely have slightly different labels to axes.

Link to git code for plotting: https://git.ligo.org/leendert.schrader/alm-beam-simulation-for-sqz/-/tree/main

Images attached to this comment
Displaying report 1-1 of 1.