10.215kHZ PI that we pass over during Thermlalization

TJ noticed the DCPD signals rising ~25 minutes into the lock, we found this was a PI t 10.215kHz. Matt notes that we've seen this at the start of each lock even before we changed the ring heaters last. It just damps down again by itself, we expect as the HOM pass over it. Plots attached. This PI doesn't appear to be in our monitored PI list.