PRC2 and CSOFT Yaw stepped out right before the LL. TMSX_Y started to oscillate 4 seconds before the LL, PM1_Y had an excursion ~3 seconds before the LL.
Images attached to this comment
ryan.crouch@LIGO.ORG - 18:09, Wednesday 23 July 2025 (85954)
Looking at all the suspicious things together, it looks like TMSX_Y started moving first before any of the ASC loops? Only ~100ms before CSOFT started moving.
PRC2 and CSOFT Yaw stepped out right before the LL. TMSX_Y started to oscillate 4 seconds before the LL, PM1_Y had an excursion ~3 seconds before the LL.
Looking at all the suspicious things together, it looks like TMSX_Y started moving first before any of the ASC loops? Only ~100ms before CSOFT started moving.