Reports until 11:47, Thursday 24 July 2025
H1 CAL
elenna.capote@LIGO.ORG - posted 11:47, Thursday 24 July 2025 (85963)
L2 Calibration line height lowered

I lowered the L2 calibration line height by a factor of 2.2- the measurement uncertainty was around 0.0007, and our target is 0.002. Now the uncertainty seems to lie around 0.0016. I updated the guardian and SDF with the new injection amplitude- 3.5.