Stepping OMC ASC loop offsets to improve optical gain

Summary: I didn't see any conclusive difference that showed we could get an improvement from a change in a particular direction. Since the optical gain moves around so much its hard to see the effect of small or quick changes. It might be worth doing a longer test were we step these offsets in sets of three different values per dof (3*4 steps) and do each for five mins (~ 1 hour).


 

I orginally was trying to use the results of our last injection (2025-06-16 20:41:15 UTC to2025-06-16 21:01:57 UTC ) to determine how to change the OMC ASC alignment.

Using /ligo/home/jennifer.wright/git/2025/OMC_Alignment/20250116_OMC_Alignment_EXC.xml we injected four low frequency lines into H1:OMC-ASC_{POS,ANG}_{X,Y}_EXC - these channels come after the filter banks (ASC-OMC_A_PIT_OFFSET, then ASC-OMC_B_PIT_OFFSET, then ASC-OMC_A_YAW_OFFSET, then ASC-OMC_B_YAW_OFFSET) where the nominal offsets are set. The injection is at four different frequencies (0.0113, 0.0107, 0.0097 and 0.0103 Hz).

 

The analysis then involves looking at the 410 Hz line height on the OMC DCPD SUM to interpret the effect on optical gain.

 

However it is not clear from this plot that the full range of phase space has been looked at - ie we have not changed the offset with a large enough amplitude to check their current position is optimal.


 

Thus today I stepped ASC-OMC_A_PIT_OFFSET, then ASC-OMC_B_PIT_OFFSET, then ASC-OMC_A_YAW_OFFSET, then ASC-OMC_B_YAW_OFFSET - these filter banks are before the ones used above in the model and are the ones where the offset for the QPDs is set. In the

Towards the last three measurements we realised that we need to do larger changes ~0.04 counts to see a change in kappa C then wait for 3-4 mins. On the long term trend none of these changes seemed to produce any meausurable gain, but it would be good to repeat this measurement with longer times spent at each step and perhaps go in slightly larger steps.

The ndscope template is in /ligo/home/jennifer.wright/git/2025/OMC_Alignment/20250724_change _offsets_by_hand.yaml

Images attached to this report
Comments related to this report
jennifer.wright@LIGO.ORG - 18:04, Tuesday 19 August 2025 (86468)

Jennie, Elenna,

 

The other day Elenna noticed some coherence between DARM and the light reflected from the OMC (OMC-REFL_A_LF_OUT_DQ) and wondered if this implies our mode-matching or OMC alignment is bad.

So I took some times from above when one of the offsets was non-nominal and we compared the coherence between these times, since it shows some small change maybe there is still some tuning we could do of these offsets to recover some optical gain.

 

The measurement is saved in /ligo/home/jennifer.wright/git/2025/OMC_Alignment/Jennie_OMC_offsets.xml

Each was around 3 minutes long apart from the starting measurement which we took from a quiet time before the measurements.

The measurement times are:

nominal settings: 16:29:21 UTC

PITCH A: 17:54:11:UTC

PITCH B: 18:39:57 UTC

YAW A :18:58:10 UTC

YAW B :19:23:20 UTC

 

Images attached to this comment