Some of the non-stationary features in DARM

As part of the DQ shift, I was looking at the DARM spectrorams to identify non-stationary (of the order of hours) features. Below is a list of features that were relatively obvious from the spectrograms. I used GDS-CALIB_STRAIN_NOLINES channel data for July, 21, 2025. In order to see the frequency variations a bit clear, i used 100-second spectrograms.

Images attached to this report
Comments related to this report
shivaraj.kandhasamy@LIGO.ORG - 03:06, Monday 28 July 2025 (86010)DetChar

To identiy possible sources for the features listed in the above alog, I have produced spectrograms of some of the important auxiallary and environmental channels. I used the same fft parameters as the GDS plots and also restricted to same frequency regions as that of GDS plots above. All these plots available at link.