We previoiusly swapped the satamp boxes for TMSX M1 F1/F2/F3/LF (85980), and I had just put in the generic 5.31:0.0969 zp compensation filters in at the time since the 'best possible' filters for that satamp were under the optic name of the OMC in the txt files. Now that we've decided that we're going to be keeping this satamp box in, we have fixed the naming in the txt files and so I was able to update the compensation filters to be the 'best possible' for each satamp channel (output). These were loaded in.