As the next step in fine-tuning the SR3 Y estimator, we needed to retake the SUSPOINT to M1 measurements as well as OLTFs so they could be used to calculate the filter modules for the suspoint and drive estimators. I took those measurements today.
General setup:
- in HEALTH_CHECK (but damping back on)
- damping for Y changed from -0.5 to -0.1
- OSEMINF gains and DAMP FM7 turned on (and left on afterwards 86070)
SUSPOINT to M1:
Data for those measurements can be found in /ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/SR3/Common/Data/2025-07-29_1730_H1ISIHAM5_ST1_WhiteNoise_SR3SusPoint_{L,T,V,R,P,Y}_0p02to50Hz.xml r12492
M1 to M1 (OLTFs):
After this, the next steps are to take regular transfer functions with the above setup of Y having -0.1 damping
That data is in /ligo/svncommon/SusSVN/sus/trunk/HLTS/H1/SR3/SAGM1/Data/2025-07-29_1830_H1SUSSR3_M1_WhiteNoise_{L,T,V,R,P,Y}_0p02to50Hz_OpenLoopGainTF.xml r12493
Reminder that these open loop transfer functions were taken with the damping Y gain of -0.1, so they should not be taken as 'nominal' OLTFs.
The M1 to M1 TFs were supposed to be regular TFs so here is the alog for those: 86202