[Jeff K. Arnaud P.]
The TF of ITMX ran over the week end from matlab 2012b, had the same issue that I encountered on friday morning. The log file saved gives the following error message :
So we investigated and fixed the problem which was basically :
Since we have higher sampling rates for the new witness channels recently implemented, (L1_WIT_$DOF_DQ / L2_WIT_$DOF_DQ), the function "get_sch_TF_M0_v9DQ_sma.m" was trying to downsample those channels to 256Hz right after getting the data out of the frame. If the function get_data failed (randomly the case), it would try to decimate an empty vector, and was returning the above error and quit the script.
We modified get_sch_TF_M0_v9DQ_sma in order to make sure the data is beeing recovered in a while loop before decimating the signal.