Displaying report 1-1 of 1.
Reports until 17:33, Monday 04 August 2025
H1 SUS (SEI, SUS)
edgard.bonilla@LIGO.ORG - posted 17:33, Monday 04 August 2025 (86187)
Created new OSEM estimator fits for SR3 Yaw (once again)

Ivey, Edgard.

Ivey made even newer fits for the M1 SR3 in "Yaw using the measurements that Oli took in [LHO: 86075].

These fits are pretty similar to the last set we got in [LHO: 85446]. However, this time, Ivey used vectfit3 for both of the fits directly. She only manually ensured that the zeros were all in the left-hand-plane, and that there were no additional zeros in the M1 to M1 transfer function, even if the measurements show a bit extra phase loss compared to the fit.

These are the fits obtained [see figures for an eye-test of the goodness of fit]:

For the ISI to M1 transfer function:
'zpk([0,0,-0.027+20.489i,-0.064+11.458i,-0.027-20.489i,-0.064-11.458i],[-0.072+6.395i,-0.072-6.395i,-0.096+14.454i,-0.096-14.454i,-0.062+21.267i,-0.062-21.267i],-0.001)'

The M1 to M1 transfer function:
'zpk([-0.002+19.246i,-0.005+8.312i,-0.002-19.246i,-0.005-8.312i],[-0.065+6.392i,-0.065-6.392i,-0.076+14.442i,-0.076-14.442i,-0.055+21.272i,-0.055-21.272i],12.085)'

The fits were added to the Sus SVN and live inside '/ligo/svncommon/SusSVN/sus/trunk/HLTS/Common/FilterDesign/Estimator/fits_H1SR3_2025-07-29.mat' .

The filters can be installed using 'make_SR3_yaw_model.m', which lives in the same folder [for reference, see LHO: 84041, where Oli got the fits running for a test]. This last script, as well as 'make_SR3_yaw_blend.m' were updated to work with the new naming convention for the estimator block.

Hopefully we will get to test the estimator again soon!

Images attached to this report
Displaying report 1-1 of 1.