IX Laser Power HD PD jumped down abruptly from 51.8W to 51.6W but is still above expected 35W. IY power is stable.
Both IX and IY Laser Power MTR is above 1.7 for both IY and IX, (guide says that's the max) but it's only 1.73ish so I assume that's fine.
IX chiller flow (purple) is not steady as the guide says is expected but jumps abruptly between 3.6 and 3.8 with 12 jumps back and forth in the last month. Additionally, the IX flowrate is also fluctuating and in one instance going below 2.8 GPM (which is the max). It seems to jump to 3 then slowly deteriorate over the course of a week and then jump back as the chiller changes it. Potentially worth investigation? Might be a known issue?
HWS FAMIS:
Both HWS Sleds have been going down for the whole month. They started at ~2mW 42 days ago but IX went down to 1.3mW and IY to 1.7mW. Judging from last month's, I believe this is normal behavior?
22 days ago, both IX and IY probe spherical powers jumped. IX went from ~60 to 120 and IY from 30 to 60.
While measured and simulated trends have a similar shape, the measured trends have that jump 22 days ago that I described above.