Displaying report 1-1 of 1.
Reports until 13:36, Saturday 09 August 2025
H1 CAL
anthony.sanchez@LIGO.ORG - posted 13:36, Saturday 09 August 2025 (86279)
Well Thermalized Calibration Sweep

Command Ran:
pydarm measure --run-headless bb
2025-08-09 12:06:24,938 config file: /ligo/groups/cal/H1/ifo/pydarm_cmd_H1.yaml


notification: new test result
notification: end of measurement
notification: end of test
diag> save /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20250809T190625Z.xml
/ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20250809T190625Z.xml saved
diag> quit
EXIT KERNEL

2025-08-09 12:11:35,388 bb measurement complete.
2025-08-09 12:11:35,388 bb output: /ligo/groups/cal/H1/measurements/PCALY2DARM_BB/PCALY2DARM_BB_20250809T190625Z.xml
2025-08-09 12:11:35,389 all measurements complete.
------------------------------------------------------------------------------------

Command Ran:
gpstime;python /ligo/groups/cal/src/simulines/simulines/simuLines.py -i /ligo/groups/cal/H1/simulines_settings/newDARM_20231221/settings_h1_20250212.ini;gpstime
Approximate Time ran:
~UTC:  2025-08-09 19:17:38
~GPS: 1438802276

2025-08-09 19:17:38,535 | INFO | File written out to: /ligo/groups/cal/H1/measurements/DARMOLG_SS/DARMOLG_SS_20250809T191738Z.hdf5
2025-08-09 19:17:38,543 | INFO | File written out to: /ligo/groups/cal/H1/measurements/PCALY2DARM_SS/PCALY2DARM_SS_20250809T191738Z.hdf5
2025-08-09 19:17:38,548 | INFO | File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L1_SS/SUSETMX_L1_SS_20250809T191738Z.hdf5
2025-08-09 19:17:38,553 | INFO | File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L2_SS/SUSETMX_L2_SS_20250809T191738Z.hdf5
2025-08-09 19:17:38,557 | INFO | File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L3_SS/SUSETMX_L3_SS_20250809T191738Z.hdf5
2025-08-09 19:17:38,557 | INFO | Overall driving parameters:
2025-08-09 19:17:38,557 | INFO | Preparing to scan the following sweeps ['DARM_OLGTF', 'PCALY2DARMTF', 'L1_SUSETMX_iEXC2DARMTF', 'L2_SUSETMX_iEXC2DARMTF', 'L3_SUSETMX_iEXC2DARMTF']
2025-08-09 19:17:38,557 | INFO | With parameters: BW: 0.14; DrivingCycles: 40; MinDriveTime: 3; Averages: 5; Overlap: 0; RampUpTime: 3; RampDownTime: 3; SettleTimePercent: 10; ScanTime: 1229
2025-08-09 19:17:38,558 | INFO | For DARM_OLGTF, prepaing EXC: H1:LSC-DARM1_EXC; A: ['H1:LSC-DARM1_IN2', 'H1:LSC-DARM1_EXC']; B: ['H1:LSC-DARM1_IN1', 'H1:LSC-DARM1_IN2']
2025-08-09 19:17:38,560 | INFO | For PCALY2DARMTF, prepaing EXC: H1:CAL-PCALY_SWEPT_SINE_EXC; A: ['H1:CAL-PCALY_RX_PD_OUT_DQ', 'H1:CAL-PCALY_RX_PD_OUT_DQ', 'H1:CAL-PCALY_TX_PD_OUT_DQ']; B: ['H1:CAL-DELTAL_EXTERNAL_DQ', 'H1:LSC-DARM_IN1_DQ', 'H1:LSC-DARM_IN1_DQ']
2025-08-09 19:17:38,562 | INFO | For L1_SUSETMX_iEXC2DARMTF, prepaing EXC: H1:SUS-ETMX_L1_CAL_EXC; A: ['H1:SUS-ETMX_L1_CAL_EXC']; B: ['H1:LSC-DARM_IN1_DQ']
2025-08-09 19:17:38,563 | INFO | For L2_SUSETMX_iEXC2DARMTF, prepaing EXC: H1:SUS-ETMX_L2_CAL_EXC; A: ['H1:SUS-ETMX_L2_CAL_EXC']; B: ['H1:LSC-DARM_IN1_DQ']
2025-08-09 19:17:38,565 | INFO | For L3_SUSETMX_iEXC2DARMTF, prepaing EXC: H1:SUS-ETMX_L3_CAL_EXC; A: ['H1:SUS-ETMX_L3_CAL_EXC']; B: ['H1:LSC-DARM_IN1_DQ']
2025-08-09 19:17:38,579 | INFO | Start data collection while monitoring 5 processes
2025-08-09 19:17:49,583 | INFO | Scanning frequency 7.97 in Scan : PCALY2DARMTF on PID: 2113968
2025-08-09 19:17:49,583 | INFO | Scanning frequency 12.86 in Scan : L1_SUSETMX_iEXC2DARMTF on PID: 2113971


...
...
...


2025-08-09 19:40:11,968 | INFO | Drive, on L1_SUSETMX_iEXC2DARMTF, at frequency: 12.33, and amplitude 15.362, is finished. GPS start and end time stamps: 1438803608, 1438803625
2025-08-09 19:40:12,277 | INFO | 0 still running.
2025-08-09 19:40:12,277 | INFO | gathering data for a few more seconds
2025-08-09 19:40:18,285 | INFO | Finished gathering data. Data ends at 1438803635.0
2025-08-09 19:40:18,396 | INFO | It is SAFE TO RETURN TO OBSERVING now, whilst data is processed.
2025-08-09 19:40:18,396 | INFO | Commencing data processing.
2025-08-09 19:40:18,396 | INFO | Ending lockloss monitor. This is either due to having completed the measurement, and this functionality being terminated; or because the whole process was aborted.
2025-08-09 19:41:00,607 | INFO | File written out to: /ligo/groups/cal/H1/measurements/DARMOLG_SS/DARMOLG_SS_20250809T191738Z.hdf5
2025-08-09 19:41:00,615 | INFO | File written out to: /ligo/groups/cal/H1/measurements/PCALY2DARM_SS/PCALY2DARM_SS_20250809T191738Z.hdf5
2025-08-09 19:41:00,620 | INFO | File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L1_SS/SUSETMX_L1_SS_20250809T191738Z.hdf5
2025-08-09 19:41:00,625 | INFO | File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L2_SS/SUSETMX_L2_SS_20250809T191738Z.hdf5
2025-08-09 19:41:00,630 | INFO | File written out to: /ligo/groups/cal/H1/measurements/SUSETMX_L3_SS/SUSETMX_L3_SS_20250809T191738Z.hdf5
PDT: 2025-08-09 12:41:00.802402 PDT
UTC: 2025-08-09 19:41:00.802402 UTC
GPS: 1438803678.802402

 

 

 



 

Images attached to this report
Non-image files attached to this report
Displaying report 1-1 of 1.