Closes FAMIS#26550, last checked 86343
Laser Status:
NPRO output power is 1.865W
AMP1 output power is 69.91W
AMP2 output power is 140.3W
NPRO watchdog is GREEN
AMP1 watchdog is GREEN
AMP2 watchdog is GREEN
PDWD watchdog is GREEN
PMC:
It has been locked 9 days, 22 hr 41 minutes
Reflected power = 23.7W
Transmitted power = 105.3W
PowerSum = 129.0W
FSS:
It has been locked for 0 days 7 hr and 57 min
TPD[V] = 0.8324V
ISS:
The diffracted power is around 4.0%
Last saturation event was 0 days 12 hours and 31 minutes ago
Possible Issues:
PMC reflected power is high