I compared the sensitivity from two segments over the last 24 hours where our range was about 157 Mpc and 162 Mpc each (60 second averages). Looks like a broad improvement from about 30-300 Hz when using the range comparison script.. The 350 Hz SQZ blrms are 0.25 dB better in the higher range time. It's hard to see a large change in the 38-60, 60-100 or 100-450 Hz range blrms. Whether you use a shorter or longer average, the difference between two times is 5 Mpc.
After the period of high range, the sensivity then dropped a few Mpc, which seems to match with the 38-60 Hz blrms getting slightly worse. I nearly had to squint to see it, so here's another range comparison plot with the high range time above, and then a low range time between 6-7 hours ago. This is about a 4 Mpc loss from 20-100 Hz.