PSL Weekly Report - Weekly FAMIS 26551

Closes FAMIS 26551. Last checked in alog 86515


Laser Status:
    NPRO output power is 1.873W
    AMP1 output power is 70.19W
    AMP2 output power is 141.3W
    NPRO watchdog is GREEN
    AMP1 watchdog is GREEN
    AMP2 watchdog is GREEN
    PDWD watchdog is GREEN

PMC:
    It has been locked 16 days, 0 hr 26 minutes
    Reflected power = 23.93W
    Transmitted power = 105.2W
    PowerSum = 129.1W

FSS:
    It has been locked for 0 days 0 hr and 39 min
    TPD[V] = 0.8058V

ISS:
    The diffracted power is around 4.2%
    Last saturation event was 0 days 0 hours and 39 minutes ago


Possible Issues:
    PMC reflected power is high